The horizontal autoclave, which adopts the way of gravity exchange to release the cold air from the chamber more completely to ensure the reliable sterilization. The control system automatically adjust the steam inlet and outlet according to chamber teperature during sterilizing. This Unit is an ideal equipment for clinics, scientific research institutes and other oraginzations to sterilize surgical instruments, Fabric glassed, and the culture media etc.
Characteristics
- Automatically controlled the sterilization process, easy to operate
- With drying function, suitable for the medical dressing drying.
- With over-temperature, over-pressure auto-protect
- The door can not be open until pressure in the chamber is reduced to 0.27MPa. And the unit can not started on if the door doesnt close well
- The safety valve will be open automatically when the inner pressure over 0.24MPa and the steam would be exhausted the water tank.
- Automatically cut off the power if the water-lack happened, and alarm at the meanwhile.
- Fully stainless steel structure.
- With printer
Technical data | WS-150YDD | WS-200YDD | WS-280YDD |
Sterilizing chamber volume | 150L φ440×1000mm |
200L φ515×1000mm |
280L φ600×1000mm |
working pressure | 0.22 MPa | ||
working temperature | 134℃ | ||
Adjustment of temperature | 40℃ – 134℃ | ||
Time of sterilization | 0-99 min | ||
Time of drying | 0-99 min | ||
Heat average | ≤± 2℃ | ||
Power | 9KW/380V 50Hz | 9KW/380V 50Hz | 12KW/380V 50Hz |
Dimension | 1400×650×1600 | 1400×750×1700 | 1520×910×1900 |
Transport dimension | 1560×820×1820 | 1560×910×1880 | 1680×1080×2100 |
G.W/N.W | 430/340 KG | 436/350 KG | 570/462 KG |