Product Introduction
Feyond series multi-mode microplate reader is specially designed for medical, biological and pharmaceutical research and development institutions to meet the needs of various drug development and life science research. The high-quality detection performance ensures high-quality analysis based on molecular biology, biochemistry and cytology.
In addition to the most basic absorbance, fluorescence and luminescence detection functions, high-performance fluorescence polarization and time-resolved fluorescence detection can also be selected. The instrument is compatible with the client-side modular upgrade function, and users can upgrade and equip with microplates and automatic injectors according to their needs.
■ UV/Vis Absorbance (ABS)
Wavelength selection is done by using an advanced monochromator system. Any wavelength between 200 to 1000 nm can be selected. Almost satisfy all absorbance detection applications.
■ Fluorescence (FL)
Fluorescence detection adopts the filter-based optics detection, which is composed of xenon lamp, filter and PMT and can best fit the characteristics of fluorescent dyes.
■ Luminescence (LUM)
Feyond series luminescence microplate readers show excellent sensitivity and wide dynamic range in both glow and flash based assays. The optimized light path minimizes signal crosstalk between holes and ensures the accuracy of experimental results.
■ Time-Resolved Fluorescence (TRF)
Time-resolved fluorescence is based on lanthanide elements as dyes. After the excitation light is turned off, the emitted light can still be continuously expressed and released, thereby eliminating the interference of excitation light and scattered light. Time-resolved fluorescence has high sensitivity, strong specificity, good stability, and short operation flow. It is suitable for ultra-micro analysis in biology and medicine, hormone detection, viral hepatitis marker detection, target cell marker detection, and drug screening.
■ Fluorescence Polarization (FP)
The optimized optical path design of Feyond-A500 combined with the performance of the fast switching polarizer can effectively reduce the detection deviation. This function is often used to detect the interaction between small molecules and macromolecules, such as the determination of drugs and hormones, tyrosine kinase detection, receptor / ligand
research, protein / polypeptide interactions, DNA / protein interactions, etc.
01:
Rotating Touch Screen
The instrument is equipped with a 10-inch touch screen, whose angle can be converted through the LCD control button to facilitate the researcher’s setting of the instrument.
02:
High-Quality Stable Performance
It has reference optical path to ensure more stable detection data. After the instrument is started, the light source, mono- chromator, detector, position, etc. will be calibrated automati- cally to ensure the stable and reliable operation of the instru- ment.
![]() |
03:
Modular Filter
The easy-to-disassemble modular filter will bring an economi- cal and highly sensitive solution to your fluorescence detec- tions. Only by scanning the QR codes on the block, the instru- ment can read the filter information to ensure accurate experi- mental parameters for convenient and quick operation.
![]() |
04:
MSS-2 Automatic Injector Block (Optional)
The automatic injector block can be purchased at the same time with the device and can also be upgraded later.
Liquid injection protection: The software has a liquid injection protection function to effectively prevent the risk of sample overflow.
![]() |
05:
Easy-to-Use and Flexible Software
1. User Authority Classification
Intuitive selection of function modes and programs and results are stored independently
2. Intuitive Interface Display
Programs and results are stored independently, making it easy to be found the required applications
3. Powerful Data Analysis and Process
Provide multiple data processing methods including blank subtraction, standard curve creation, qualitative analysis, quality control, kinetics and spectral analysis
4. FTP (File Transfer Protocol)
Upload the data directly to a computer with a FTP server, and users can view the data results at any time in the authorized folder
![]() |
06:
u-Nano Ultra-Micro Plate (Optional)
1. Quickly complete high-throughput quantification of nucleic acid and proteins without samples dilution
2. The independent software can quickly give the sample concen- tration and purity report
3. 1~16 samples can be detected at the same time, only 2~4 μL sample volume is needed
![]() |
07:
ReaderIt-ll PC Analysis Software (Optional)
1. Can provide a more comprehensive and complex data analysis algorithm than the instrument APP software
2. Data export is convenient and fast. Detailed result reports can be created through built-in tools
![]() |
Spec:
Model | Feyond-A300 | Feyond-A400 | Feyond-A500 | |
Absorbance | Light source | Xenon lamp | ||
Detector | PD | |||
Wavelength accuracy | 2 nm | |||
Wavelength repeatability | 0.2nm | |||
Half width | <2.5 nm | |||
Wavelength range | 200-1000 nm, 1 nm step | |||
Measure range | 0-4 OD | |||
Resolution | 0.0001 OD | |||
Accuracy @450nm | 96-precision mode: ± (1.0 %+0.003 Abs) @(0.0-2.0 Abs] ±2.0 % @(2.0-3.0 Abs] |
|||
Repeatability @450nm | CV < 1.0 % or SD < 0.003 fast (0.0-3.0Abs] | |||
CV < 0.5 % or SD < 0.003 accurate (0.0-3.0Abs] | ||||
Stray light | 0.1% @220 nm | |||
Linear @450nm | R2 ≥ 0.999 @[0.0 – 3.0 Abs] | |||
Reading time | 96-well plate: fast <15 s | |||
Luminescence | Detector | PMT | ||
Detection limit | 15 amol/hole; 5 amol/hole (photon PMT) | |||
Linear dynamic range | 6 logs | |||
Crosstalk | ≤ 0.005 % | |||
Wavelenth range | 200-850 nm | |||
Fluorescence | Reading mode | Top reading | ||
Excitation light source | Xenon lamp | |||
Detector | PMT | |||
Wavelength range | EX: 200-1000 nm; EM: 270-850 nm | |||
Filter EX/EM | 3 groups: EX470/EM525, EX523/EM564, EX624/EM692 (other wavelengths can be replaced) | |||
Detection limit | 1 pM | |||
Linear dynamic range | 6 logs | |||
TRF | Wavelength range | – | EX: 200-1000 nm; EM: 270-850 nm | |
Detection limit | – | 0.05 pM | ||
FP | Wavelength range | – | – | λ1-λ2 |
Detection limit | – | – | 5 mP |
Basic Spec:
Model | Feyond-A300 | Feyond-A400 | Feyond-A500 |
Plate | 6-384 well | ||
Accessories | Microplate, injector | ||
Shaking mode | Linear, annular, double annular | ||
Incubation temperature | RT+4 °C ~ 45 °C | ||
Temperature uniformity | ±0.5 °C @37 °C | ||
Software interface | Chinese / English | ||
Screen size | 10-inch | ||
Operation method | Capacitive screen touch, mouse | ||
Data capacity | 10 GB | ||
Compatibility | Support PC software, Win7/Win10 64 bit | ||
Network transmission | The test data report can be uploaded to the PC server through FTP | ||
Instrument port | 2 USB Type A ports, 1 USB Type B port, 1 Ethernet port, Rs232 bus interface (connected to the injector) | ||
Power supply | AC 100-240 V, 50-60 Hz | ||
Size (W×D×H) | 420×550×386 mm | ||
Weight | 33 kg |
Microplate
|
|
Sample number
|
1-16
|
Sample detection volume
|
2-4 μL
|
Automatic Injector
|
|
Quantity
|
1/2
|
Dispensing volume
|
5-1000 μL, 1 μL increment
|
Liquid injection speed
|
125-500 μL/s
|
Accuracy
|
±1 μL @5-50 μL
±2 % @51-1000 μL |
Waste liquid collection
|
50 mL
|
Software
|
|
Analysis software
|
ReaderIt-II software
|
Code
|
Product Description
|
AS-19050-00
|
Feyond-A300 multi-mode microplate reader
|
AS-19060-00
|
Feyond-A400 multi-mode microplate reader
|
AS-19070-00
|
Feyond-A500 multi-mode microplate reader
|
AS-19051-01
|
Lum-check standardization light microplate
|
AS-19051-02
|
470-525 fluorescence filter (standard)
|
AS-19051-03
|
523-564 fluorescence filter (standard)
|
AS-19051-04
|
624-692 fluorescence filter (standard)
|
AS-19051-05
|
365-612 time-resolved fluorescence filter (standard)
|
AS-19051-06
|
485-530 fluorescence polaroid filter (standard)
|
AS-19051-41
|
LUM-470 filter (standard)
|